262
Chapter 5 Error Messages
Execution Errors
5
-222 Data out of range;
period; value clipped to ...
This generic message indicates that the waveform period has been
limited to an upper or lower boundary.
-222 Data out of range;
frequency; value clipped to ...
This generic message indicates that the waveform frequency has been
limited to an upper or lower boundary.
-222 Data out of range;
user frequency; value clipped to upper limit
This generic message indicates that the waveform frequency has been
limited to an upper boundary due to the selection of the arbitrary
waveform function (APPL:USER or FUNC:USER command).
-222 Data out of range;
ramp frequency; value clipped to upper limit
This generic message indicates that the waveform frequency has been
limited to an upper boundary due to the selection of the ramp waveform
function (APPL:RAMP or FUNC:RAMP command).
-222 Data out of range;
pulse frequency; value clipped to upper limit
This generic message indicates that the waveform frequency has been
limited to an upper boundary due to the selection of the pulse waveform
function (APPL:PULS or FUNC:PULS command).
-222 Data out of range;
burst period; value clipped to ...
This generic message indicates that the burst period has been limited to
an upper or lower boundary.
-222 Data out of range;
burst count; value clipped to ...
This generic message indicates that the burst count has been limited to
an upper or lower boundary.