265
Chapter 5 Error Messages
Execution Errors
4
5
-222 Data out of range;
duty cycle; value clipped to ...
The duty cycle is limited to values between 20% and 80% when the
frequency is less than 25 MHz.
Duty Cycle: 20% to 80% (frequency <
25 MHz)
40% to 60% (25 MHz < frequency <
50 MHz)
50% (frequency > 50 MHz)
-222 Data out of range;
duty cycle limited by frequency; value clipped to upper limit
The duty cycle is limited to 50% when the frequency is greater
than 50 MHz.
Duty Cycle: 20% to 80% (frequency <
25 MHz)
40% to 60% (25 MHz < frequency <
50 MHz)
50% (frequency > 50 MHz)
-313 Calibration memory lost;
memory corruption detected
The non-volatile memory used to store the function generator’s calibration
constants has detected a checksum error. This error may be the result of
a device failure or extreme conditions such as lightning or strong
magnetic fields.
-314 Save/recall memory lost;
memory corruption detected
The non-volatile memory used to store instrument states has detected a
checksum error. This error may be the result of a device failure or
extreme conditions such as lightning or strong magnetic fields.
-315 Configuration memory lost;
memory corruption detected
The non-volatile memory used to store the function generator’s
configuration settings (i.e., remote interface settings) has detected a
checksum error. This error may be the result of a device failure or
extreme conditions such as lightning or strong magnetic fields.