Chapter 5 Error Messages
Execution Errors
5
-350 Queue overflow
The error queue is full because more than 20 errors have occurred.
No additional errors are stored until you remove errors from the queue.
The error queue is cleared by the *CLS (clear status) command or when
the power is cycled. The errors are also cleared when you read
the queue.
The error queue
is not
cleared by an instrument reset
(*RST command).
-361 Parity error in program message
This error is most likely to occur when the function generator’s parity
setting (RS-232 interface) does not match the computer’s parity setting.
This error may also occur in the presence of noise on the RS-232 cable.
-362 Framing error in program message
This error is most likely to occur when the number of stop bits for the
computer (RS-232 interface) does not match the function generator’s
setting (fixed at 1 stop bit).
-363 Input buffer overrun
Too many characters have been sent to the function generator while
configured to use the RS-232 interface. This error typically occurs when
you have selected no handshaking of data between the computer and the
function generator. To avoid this error, select one of the handshake
modes available for the 33250A (see “Remote Interface Configuration”
on page 118 for more information).

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish

Related product manuals