General Purpose Switch Modules 6
34980A User’s Guide 191
Reading Cycle Count and Resetting Modules to Power-On State
Example: Reading the cycle count for a relay (all switch modules) The
following command returns the relay cycle count on channel 7 and
channel 16 for a module in slot 1.
DIAGnostic:RELay:CYCLes? (@1007,1016)
Example: Clearing the cycle count for a relay (all switch modules) The
following command resets the relay cycle count on channels 7 and 16 for
a module in slot 1.
DIAGnostic:RELay:CYCLes:CLEar (@1007,1016)
Example: Resetting Module(s) to power-on state (all modules) The following
command resets a module in slot 4 to its power- on state.
SYSTem:CPON 4