Microwave Switch/Attenuator Driver 8
34980A User’s Guide 215
Using Single Drive Switches and Attenuators
Some microwave switches require a single drive. With single drive devices
the channel numbering is not consecutive across all channels in a bank
(refer to the channel numbering description on page 212).
The 34945A can provide single drive devices with either pulsed or
continuous drive current. Settings and parameters for continuous drive
mode are given in the next section.
Continuous Drive Current
Driving non- latching devices requires a power supply capable of handling
sustained high current requirements. You may only use continuous drive
current with channel configured for single drive. Additionally, to prevent
power supply loading, care must be taken when operating more than one
continuous drive at a time. The actual drive may be configured as either
TTL or open- collector operation.
Using Continuous Drive
The diagram below illustrates the continuous drive signals for two
channels (switches) and the relationship of the drive parameters to the
power supply requirements.
As shown in the diagram, the drive signal is initially applied to channel 1.
Drive is applied to channel 2 only after a power supply recovery period
has elapsed T(Recovery). The power supply recovery time is set using the
the ROUTe:CHANnel:DRIVe:TIMe:RECovery command. This parameter may
be set individually for each channel or will default to 0.0 ms following
either a SYSTem:RMODule:RESet or ROUTe:RMODule:BANK:PRESet
command.
If you are verifying the channel closure (see page 218), you may also
specify a T(Settle) parameter. This parameter ensures the switch has had
time to change state before the position indicator is evaluated. This
parameter may be set individually for each channel or will default to
0.0 ms following either a
SYSTem:RMODule:PRESet or
ROUTe:RMODule:BANK:PRESet command.
T(Setttle)
T(Recovery)
Drive Ch 1
Drive Ch2
Start Drive
Channel 1
Start Drive
Channel 2
Channel 1 Position
Indicators Evaluated

Table of Contents

Other manuals for Agilent Technologies 34980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34980A and is the answer not in the manual?

Agilent Technologies 34980A Specifications

General IconGeneral
BrandAgilent Technologies
Model34980A
CategorySwitch
LanguageEnglish

Related product manuals