54 34980A User’s Guide
2 Features and Functions
Sample Count
The sample count sets the number of auto- triggered samples the internal
DMM will take per channel per trigger. The sample count applies to both
scanning and stand- alone DMM measurements (with no scan list). The
front- panel sample annunciator (“*) turns on during each measurement.
Sample count for Stand-Alone DMM Mode
Sample count for Scanning Mode
t
Sample Count
(1 to 500,000 samples)
Trigger Trigger
t
Sweep 1Trigger Sweep 2
Sweep n
Trigger
. . .
Sweep Count
Ch 1 Ch 3 Ch 4Ch 2 Ch 5 Ch 6
Sample Count
(1 to 500,000 samples)