EasyManua.ls Logo

Agilent Technologies 34980A

Agilent Technologies 34980A
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42 34980A User’s Guide
2 Features and Functions
Front Panel Operation:
DMM or Channel (Configure) > SCALING > GAIN|OFFSET|UNITS
To define the label on the selected channel, press the arrow keys to move
the cursor to a specific position and then turn the knob to select the
desired letter or number. To clear the label on the selected channel,
change each character to “ ^ ” (starting with the rightmost character) and
then press the left arrow key to move to the next character.
Remote Interface Operation: Use the following commands to set the gain,
offset, and custom measurement label.
CALC:SCALE:GAIN 1.2,(@1003)
CALC:SCALE:OFFSET 10,(@1003)
CALC:SCALE:UNIT 'PSI',(@1003)
After setting the gain and offset values, send the following command to
enable the scaling function on the specified channel.
CALC:SCALE:STATE ON,(@1003)

Table of Contents

Other manuals for Agilent Technologies 34980A

Related product manuals