58 34980A User’s Guide
2 Features and Functions
AC Voltage, AC Current (for all ranges):
Frequency, Period:
Digital Input, Totalize:
Front Panel Operation: Channel (Configure) > CHANNEL DELAY > AUTO
Once you have added the specified channel to the scan list, the channel
delay choice will be visible in the menu.
Interface Operation: The following command enables an automatic channel
delay on the specified channels.
ROUT:CHAN:DELAY:AUTO ON,(@1003,1013)
Selecting a specific channel delay using the ROUTe:CHANnel:DELay
command (see Channel Delay” on page 56) disables the automatic
channel delay.
AC Filter Channel Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
7.0 seconds
1.0 second
120 ms
AC Filter Channel Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
600 ms
300 ms
100 ms
Channel Delay
0 seconds