EasyManua.ls Logo

Agilent Technologies 35670A - To arrange traces on the screen

Agilent Technologies 35670A
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-Weight Filter
Measurement Lower
Limit
(dB)
Upper Limit
(dB)
Measured Value
(dB)
Pass/Fail
Ch 1, 10 Hz -5 2
Ch 2, 10 Hz -5 2
Ch 1, 31.62 Hz -1 1
Ch 2, 31.62 Hz -1 1
Ch 1, 100 Hz -0.7 0.7
Ch 2, 100 Hz -0.7 0.7
Ch 1, 1000 Hz -0.7 0.7
Ch 2, 1000 Hz -0.7 0.7
Ch 1, 10000 Hz -3 2
Ch 2, 10000 Hz -3 2
Ch 1, 25120 Hz -4.5 2.4
Ch 2, 25120 Hz -4.5 2.4
Channel Match
Measurement Lower
Limit
Upper Limit Measured Value Pass/Fail
Two Ch, 2/1, 7 dBV FS Mag -0.04 dB 0.04 dB dB
Two Ch, 2/1, 7 dBV FS Phs -0.5 deg 0.5 deg deg
Two Ch, 2/1, -13 dBV FS Mag -0.04 dB 0.04 dB dB
Two Ch, 2/1, -13 dBV FS Phs -0.5 deg 0.5 deg deg
Two Ch, 2/1, -33 dBV FS Mag -0.04 dB 0.04 dB dB
Two Ch, 2/1, -33 dBV FS Phs -0.5 deg 0.5 deg deg
Two Ch, 2/1, 7 dBV -20dBfs Mag -0.08 dB 0.08 dB dB
Two Ch, 2/1, 7 dBV -20dBfs Phs -0.5 deg 0.5 deg deg
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Verifying Specifications Agilent 35670A
Operation Verification Test Record - Two Channel
8 of 10