A-Weight Filter
Measurement Lower
Limit
(dB)
Upper Limit
(dB)
Measured Value
(dB)
Pass/Fail
Ch 1, 10 Hz -5 2
Ch 2, 10 Hz -5 2
Ch 1, 31.62 Hz -1 1
Ch 2, 31.62 Hz -1 1
Ch 1, 100 Hz -0.7 0.7
Ch 2, 100 Hz -0.7 0.7
Ch 1, 1000 Hz -0.7 0.7
Ch 2, 1000 Hz -0.7 0.7
Ch 1, 10000 Hz -3 2
Ch 2, 10000 Hz -3 2
Ch 1, 25120 Hz -4.5 2.4
Ch 2, 25120 Hz -4.5 2.4
Channel Match
Measurement Lower
Limit
Upper Limit Measured Value Pass/Fail
Two Ch, 2/1, 7 dBV FS Mag -0.04 dB 0.04 dB dB
Two Ch, 2/1, 7 dBV FS Phs -0.5 deg 0.5 deg deg
Two Ch, 2/1, -13 dBV FS Mag -0.04 dB 0.04 dB dB
Two Ch, 2/1, -13 dBV FS Phs -0.5 deg 0.5 deg deg
Two Ch, 2/1, -33 dBV FS Mag -0.04 dB 0.04 dB dB
Two Ch, 2/1, -33 dBV FS Phs -0.5 deg 0.5 deg deg
Two Ch, 2/1, 7 dBV -20dBfs Mag -0.08 dB 0.08 dB dB
Two Ch, 2/1, 7 dBV -20dBfs Phs -0.5 deg 0.5 deg deg
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Verifying Specifications Agilent 35670A
Operation Verification Test Record - Two Channel
8 of 10

Table of Contents

Other manuals for Agilent Technologies 35670A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 35670A and is the answer not in the manual?

Agilent Technologies 35670A Specifications

General IconGeneral
BrandAgilent Technologies
Model35670A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals