EasyManuals Logo
Home>Agilent Technologies>Security Sensors>5977C Series

Agilent Technologies 5977C Series User Manual

Agilent Technologies 5977C Series
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
4 Operating in EI Mode
To Verify EI System Performance
104 5977C GC/MSD Operating Manual
To Verify EI System Performance
Materials needed for an EI HES
10 fg/µL OFN Checkout standard 3 × 1 mL (5190-0585)
Materials needed for an EI XTR, SS, or Inert source
1 pg/µL OFN EI Checkout standard 3 × 1 mL (5188-5348)
Verify the tune performance
1 Verify that the system has been pumping down for at least 60 minutes.
2 Set the GC oven temperature to 150 °C and the column flow to 1.0 mL/min.
3 In the Instrument Control view, select Checkout Tune from the Checkout
menu. The software will perform an autotune and print the report.
4 When the autotune has completed, save the method, and then select
Evaluate Tune from the Checkout menu.
The software will evaluate the last autotune, and print a System Verification
– Tune report.
Verify the sensitivity performance
1 Set up to inject 1 µL of the checkout sample shown in the Materials needed
for your source type above, either with the ALS or manually.
2 In the Instrument Control view, select Sensitivity Check from the Checkout
menu. The system displays an Alert dialog reminder about resolving the
OFN_SN method and placing the OFN sample in vial 1 when an ALS is
configured.
3 If necessary, resolve your hardware with this method and place the sample in
the vial 1 position.
4 Click OK to run the method.
When the method is completed, an evaluation report will be printed.
Verify that rms signal-to-noise ratio meets the published specification. See the
Agilent Web site at www.agilent.com/chem for specifications.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Series and is the answer not in the manual?

Agilent Technologies 5977C Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals