6 General Maintenance
To Install the EI HES Filament
5977C GC/MSD Operating Manual 167
To Install the EI HES Filament
Materials needed
Filament assembly, high efficiency dual filament (G7001-60001)
Hex ball driver, 1.5 mm (8710-1570)
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
T6 Torx driver, (8710-2548)
Procedure
1 Remove the old filament. (See “To Remove the EI HES Filaments” on
page 165.)
2 Insert the 3 pins on the dual filament through the rear of the filament block.
(See Figure 53 on page 165.)
3 Place the filament block in the source mount.
4 Use a T6 Torx screwdriver to tighten the screw securing the filament block to
the source mount.
5 Reinstall the ion source. (See “To Install the EI HES” on page 168.)
6 Close the analyzer chamber. (See “To Close the Analyzer Chamber” on
page 202.)
7 Pumpdown the MSD. (See “To Pumpdown the MSD” on page 112.)
8 Autotune the MSD.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Series and is the answer not in the manual?

Agilent Technologies 5977C Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals