3 Installing Intuvo 9000 GC Columns
To Replace an Intuvo GC Column
5977C GC/MSD Operating Manual 59
To Replace an Intuvo GC Column
This procedure applies to GCs with a single column. For a 2 column replacement
see the Agilent Intuvo 9000 Gas Chromatograph Maintaining Your GC manual.
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Tweezers (8710-2460)
Intuvo gasket, polyimide 5/pk, for temperatures < 350 °C (5190-9072)
Intuvo gasket, nickel 5/pk, for temperatures 350 °C to 450 °C (5190-9073)
Intuvo torque driver stored on oven door (5190-9571)
Procedure
1 Prepare the GC for maintenance.
From the GC panel select Maintenance > Column > Perform Maintenance >
Install Column > Start Maintenance. This procedure cools the inlet, detector,
column, Guard Chip, and other components in the flow path heated zones to
< 40 °C and configures the GC. Follow the on-screen GC prompts.
2 From MassHunter Data Acquisition, Vent the MSD (See “To Vent the
MSD” on page 109.)
3 If you are using hydrogen or other flammable gas as a carrier gas, close the
manual gas supply valve to the instrument before turning off the MSD power.
WARNING
The inlet, detector, bus components, and column can be hot enough to cause
burns. Cool heated zones to a safe handling temperature before continuing.
CAUTION
Always wear clean gloves while handling any parts that are in the flow path
going to the analyzer or parts that are located inside the analyzer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Series and is the answer not in the manual?

Agilent Technologies 5977C Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals