EasyManua.ls Logo

Agilent Technologies 5977C Series - To Install the EI HES Radiator

Agilent Technologies 5977C Series
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7CI Maintenance
To Install the EI HES Radiator
214 5977C GC/MSD Operating Manual
To Install the EI HES Radiator
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
T10 Torx screwdriver (5182-3466)
Tweezers (8710-2460)
Procedure
1 Place the radiator over the guide pins on the analyzer support and use a T10
Torx screwdriver to secure it using the (2) M3 x 12 gold plated screws (part
number G7002-20110).
2 Use tweezers to connect the wires from the EI HES radiator. (See“To
Connect/Disconnect Wiring to the EI HES Radiator” on page 207.)
WARNING
The analyzers, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
CAUTION
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.

Table of Contents

Related product manuals