EasyManua.ls Logo

Agilent Technologies 5977C Series - To Connect;Disconnect Wiring to the EI HES Radiator

Agilent Technologies 5977C Series
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7CI Maintenance
To Connect/Disconnect Wiring to the EI HES Radiator
5977C GC/MSD Operating Manual 207
To Connect/Disconnect Wiring to the EI HES
Radiator
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Pliers, long-nose (8710-1094)
Tweezers (8710-2460)
Procedure
1 Use tweezers or needle nose pliers to connect/disconnect the green ground
wire and the 5 lens wires from the radiator. Do not bend the wires any more
than necessary. (See Figure 68.)
2 Use tweezers or needle nose pliers to connect/disconnect the 2 purple
source heater wires and the 2 gray RTD wires from the ceramic source
board.

Table of Contents

Related product manuals