Page #207 background image
7CI Maintenance
To Connect/Disconnect Wiring to the EI HES Radiator
5977C GC/MSD Operating Manual 207
To Connect/Disconnect Wiring to the EI HES
Radiator
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Pliers, long-nose (8710-1094)
Tweezers (8710-2460)
Procedure
1 Use tweezers or needle nose pliers to connect/disconnect the green ground
wire and the 5 lens wires from the radiator. Do not bend the wires any more
than necessary. (See Figure 68.)
2 Use tweezers or needle nose pliers to connect/disconnect the 2 purple
source heater wires and the 2 gray RTD wires from the ceramic source
board.