2 Installing 8890 GC Columns
To Install a Capillary Column in the GC/MSD Interface Using a Standard Column Nut
5977C GC/MSD Operating Manual 47
To Install a Capillary Column in the GC/MSD
Interface Using a Standard Column Nut
This procedure is for the installation of a capillary column directly into the
analyzer. There are two types of column nuts that can be used in the GC/MSD
interface: The standard column nut explained here, and the self tightening
column nut explained in “To Install a Capillary Column in the GC/MSD Interface
Using the Self-Tightening Column Nut” on page 42.
Materials needed
Interface tip seal (G3870-20542) (See Figure 5 on page 44 and Figure 6 on
page 44.) (Not used with EI SS or Inert source)
Knurled tip seal retainer (G3870-20547) (Not used with EI SS or Inert source)
Tip seal spring (G7005-20024) (Not used with SS or Inert EI source)
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Flashlight
Magnifying loupe
Gloves, clean
Large (8650-0030)
Small (8650-0029)
Interface column nut (05988-20066)
Ferrules
0.3 mm id, for 0.10 mm id columns (5062-3507)
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
0.5 mm id, for 0.32 mm id columns (5062-3506)
0.8 mm id, for 0.53 mm id columns (5062-3512)
Septum (may be old, used inlet septum)
Safety glasses
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Series and is the answer not in the manual?

Agilent Technologies 5977C Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals