7CI Maintenance
To Remove the CI Source Radiator
212 5977C GC/MSD Operating Manual
To Remove the CI Source Radiator
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Tweezers (8710-2460)
T10 Torx screwdriver (5182-3466)
Procedure
WARNING
The analyzers, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
CAUTION
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
CAUTION
When disconnecting leads, pull on the connectors, not on the wires.