EasyManua.ls Logo

Agilent Technologies 5977C Series - To Disassemble the CI Source

Agilent Technologies 5977C Series
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7CI Maintenance
To Disassemble the CI Source
220 5977C GC/MSD Operating Manual
To Disassemble the CI Source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Nut driver, 5.5 mm (8710-1220)
Tweezers (8710-2460)
Procedure
Refer to the exploded parts view Figure 73 and CI source parts list Table 26 on
page 221 while using this procedure.
1 Remove the CI source. (See“To Remove the CI Source” on page 217.)
2 Remove the filaments. (See“To Remove the CI Source Filament” on
page 229.)
3 Separate the source heater assembly from the source body. The source
heater assembly includes the source heater, repeller, and related parts.
4 Disassemble the repeller assembly by removing the ceramic insulator from
the repeller.
5 Remove the setscrew securing the lenses to the source body.
6 Pull the lenses out of the source body and separate the lens insulator, ion
focus lens, drawout cylinder, drawout lens, and entrance lens.

Table of Contents

Related product manuals