4 Operating in EI Mode
To Set the GC/MSD Interface Temperature from MassHunter
5977C GC/MSD Operating Manual 97
4 Click Apply to download setpoints, or click OK to download setpoints and
close the window.
5 To make the new settings part of the method, select Save from the Method menu.
CAUTION
Ensure that the carrier gas is turned on and the column has been purged of air
before heating the GC/MSD interface or the GC oven.
When setting the GC/MSD interface temperature, never exceed the maximum
for your column.