Agilent 708-DS Dissolution Apparatus
Training Manual
Revision A September 2010
P/N: xx-xxxx
- 38 -
If the instrument is configured with a printer, press anywhere inside the Profile
Print Interval box to enter an appropriate frequency (in HHH:MM:SS) to print the
current test conditions (e.g., RPM, Bath Temperature). Press Ok when finished.
NOTE: A maximum of 100 ‘events’ may be stored and printed during a method.
Press Next. The third Method Properties screen displays. If applicable, place a
check in the Enable Manifold box to set the desired raising / lowering times of
the sample manifold.
Figure 6-3. Method Properties Screen 3
Place a check in the Initial Temp box if the apparatus contains vessel
temperature probes and it is desired to measure and record initial test
temperatures in all vessels. Place a check in the Final Temp box if the unit is
equipped with vessel temperature probes and it is desired to measure and record
final test temperatures in all vessels.
If applicable, press anywhere inside the Manifold Down Lead Time box and
enter an appropriate time (in MM:SS) to lower the manifold before each sample
time point. Press Ok when finished. If applicable, press anywhere inside the
Manifold Down Duration box and enter an appropriate time (in MM:SS) to keep
the manifold at the sampling position after each sample time point. Press OK
when finished.
NOTE: The sum of the Manifold Down Lead Time and Manifold Down Duration must
be ≥ 30 seconds to ensure stable vessel temperature measurements.
Place a check in the Enable Sample Point Alarm box to activate visual and
audible notifications before each sample time point. If applicable, press anywhere
inside the Alarm Lead Time box and enter an appropriate time (in MM:SS)
before each sample time point to notify the user of the upcoming event. Press Ok
when finished.
Press Next. The fourth and final Method Properties screen displays.

Table of Contents

Other manuals for Agilent Technologies 708-DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 708-DS and is the answer not in the manual?

Agilent Technologies 708-DS Specifications

General IconGeneral
BrandAgilent Technologies
Model708-DS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals