Agilent 708-DS Dissolution Apparatus
Training Manual
Revision A September 2010
P/N: xx-xxxx
- 39 -
Figure 6-4. Method Properties Screen 4
Press the button that corresponds to the appropriate sample time point to be
programmed. The Timepoint Properties screen displays.
Press anywhere inside the Timepoint box and enter an appropriate time (in
HHH:MM:SS) for the sample time point. Press Ok when finished.
Press anywhere inside the Spindle RPM box and enter an appropriate value for
the spindle speed (in RPM). Press Ok when finished. The value entered will
cause the RPM to change at the time point and continue at that rate until the next
time point is reached.
Press anywhere inside the Media
Volume box and enter an
appropriate value for the media
volume (in mL). This value
determines the depth of the manifold
at each time point. Press Ok when
finished.
Press Ok to return to the fourth
Method Properties screen.
Figure 6-5. Timepoint Properties Screen
Repeat the previous steps to setup all remaining sample time points.
Press Save to store the Method. The Select Method screen displays. Press
Return to return to the Menu screen.
To print the Method, press Reports > Print Method.... Place a check in the
Enable Printer box, if necessary. Press a number (1-5) corresponding to the

Table of Contents

Other manuals for Agilent Technologies 708-DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 708-DS and is the answer not in the manual?

Agilent Technologies 708-DS Specifications

General IconGeneral
BrandAgilent Technologies
Model708-DS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals