To pause the test for a media change (optional), press Stop > Pause Method.
When ready to continue the test, press Resume Method. To stop the test press
Stop > Stop Method.
At the conclusion of the test, a Method Complete notification displays. Press Ok
to return to the Main screen.
6.3. 708-DS Method Setup and Execution (Manual)
This section discusses how to set up and store a Manual dissolution method
within the firmware. A single dissolution method may be stored in the memory of
the instrument in this manner. To begin, from the Main screen of the 708-DS,
press Run. The Select Start Type screen displays. Press Manual. The Manual
Operation screen displays.
Figure 6-7. Manual Operation Screen
If the AutoTemp option is configured press anywhere inside the Vessel Temp
box and enter the appropriate vessel temperature (in °C) for the Method. Press
Ok when finished.
Press anywhere inside the Bath Temp box and enter the appropriate bath
temperature (in °C) for the Method. Press Ok when finished.
Press anywhere inside the Spindle Speed box and enter an appropriate value
(in RPM) for the Method. Press Ok when finished.
Press anywhere inside the Media Volume box and enter an appropriate value (in
mL) for the Method. Press Ok when finished.
Press anywhere inside the Duration box and enter the minimum time for the
Method (in HHH:MM:SS). Press Ok when finished.