Agilent 708-DS Dissolution Apparatus
Training Manual
Revision A September 2010
P/N: xx-xxxx
stored location of the Method to be printed. Press Ok > Main to return to the
Main screen.
6.2. 708-DS Test Execution (Stored Method)
This section discusses how to execute a stored dissolution method. To begin,
from the Main screen of the instrument, press Run. The Select Start Type screen
displays. Press Method. The Select Method screen displays. Press a number (1-
5) corresponding to the stored location of the Method to be executed. The Start
Options screen displays.
Figure 6-6. Test Start Screen
If applicable, press anywhere inside the Lot, Batch, and Note boxes to enter any
supporting data for the Method to be executed.
Select the Start Type to be used (Instant, Bath Temperature, or Vessel
Temperature) by pressing the appropriate circle. A Vessel Temperature start
requires the AutoTemp option. The manifold will lower and begin the test when
all vessels have attained the desired temperature. A Bath Temperature start will
cause the apparatus to begin the method when the water bath selected start
temperature is reached.
Place a check in the Time Delayed box if it is desired to delay the test start to a
specific time. If applicable, press anywhere inside the Time Delayed rectangle to
enter the desired date and time to begin the test. Press Ok when finished.
Press Ok from the Start Options screen to begin the Method. Ensure the
appropriate pre-test measurements occur, if applicable, and the Stop button
appears in place of the Run button once the test begins.