Chapter 3 205
Backing Up and Reloading Correction Constants
Analyzer Initialization
If the analyzer is an 8591C, the analyzer's startup routine will identify
it as an 8591E. Use this procedure to change the identity back to an
8591C.
Press the following keys.
PRESET (wait until preset is complete)
DISPLAY, Change Title
Use the softkeys to type in the following remote command. Don't forget
to include the semicolon (;).
FACTSET 11023,1;
NOTE A remote controller may be used in place of the execute title function.
DISPLAY, Hold
Press the following keys.
CAL, More 1 of 4, More 2 of 4, Service Cal, EXECUTE TITLE
PRESET (wait until preset is complete)
CONFIG, More 1 of 3, SHOW OPTIONS
Confirm that the correct analyzer model number is displayed.
Set the default configuration
Set the default configuration by pressing the following analyzer keys.
CONFIG, More 1 of 3, DEFAULT CONFIG, DEFAULT CONFIG