206 Chapter3
Backing Up and Reloading Correction Constants
Analyzer Initialization
Reset the power-on units
Set the power-on units by pressing the following analyzer keys.
PRESET
FREQUENCY, 2001, Hz
AMPLITUDE, More 1 of 2
INPUT Z 50 75 (so that 50 is underlined)
75 input only: Press
INPUT Z 50 75, so that 75 is underlined.
AMPLITUDE, SCALE LOG LIN (LOG), More 1 of 2
Amptd Units dBm
75 input only: Press dBmV.
AMPLITUDE, SCALE LOG LIN (LIN), More 1 of 2
Amptd Units Volts
CAL, More 1 of 4, More 2 of 4, Service Cal, STOR PWR ON UNITS