EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8590L

Agilent Technologies 8590L User Manual

Agilent Technologies 8590L
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #58 background imageLoading...
Page #58 background image
Chapter 2 59
Making Adjustments
6. Cal Attenuator Error
Procedure
The accuracy of the amplitude control attenuator is critical to the
proper calibration of the instrument; therefore, this procedure must be
carefully and accurately performed.
1. Turn the analyzer
LINE switch to OFF. Remove the instrument cover
assembly.
2. Turn the analyzer LINE switch to ON.
3. Connect the 50 output of the synthesizer/level generator to the
analyzer INPUT 50 . See Figure 2-10.
75 input: Connect the 75 output of the synthesizer/level
generator to the analyzer INPUT 75 .
Figure 2-10 Cal Attenuator Error Correction Setup
CAUTION Use only 75 cables, connectors, or adapters on instruments equipped
with 75 inputs or damage to the input connectors will occur.
4. Set the frequency of the synthesizer/level generator to 25 MHz and
the output to 19 dBm.
5. To turn the amplitude attenuator correction constants off, press the
following analyzer keys.
PRESET
CAL, More 1 of 4
CORRECT ON OFF (OFF)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8590L and is the answer not in the manual?

Agilent Technologies 8590L Specifications

General IconGeneral
BrandAgilent Technologies
Model8590L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals