98 Chapter2
Making Adjustments
15. Frequency Response of the 8590L, 8591C, and 8591E
7. On the synthesized sweeper, press CW and STEP UP, to step
through the remaining frequencies listed in Table 2-6.
At each new frequency repeat step 5 and step 6, and enter each
power sensor cal factor into the respective power meter.
Procedure
1. Zero and calibrate the measuring receiver and 1 MHz to 1.8 GHz
power sensor in log mode as described in the measuring receiver
operation manual.
2. Connect the equipment as shown in Figure 2-28.
Figure 2-28 Frequency Response Setup
3. Press INSTRUMENT PRESET on the synthesized sweeper. Set the
synthesized sweeper controls as follows:
CW ................................................................ 300MHz
FREQ STEP ................................................... 37MHz
POWER LEVEL ............................................. 9dBm

Table of Contents