6-10
Supplement for the Programmer’s Guide (Firmware Revision 7.68)
Alphabetical Command Reference
ECALMODINF
Syntax
ECALMODINF;
Description
Front Panel Equivalents
Command Description Range Response
ECALMODINF This command returns string variables on
the following information on the selected
ECal module.
Model Number: xxxxx-xxxxx
Serial Number: xxxxx
Connector Type: connector type
Last Certification: ddmmmyear
Module Number of Points: xxx
Module Start Frequency: xxxxx
a
Module Stop Frequency: xxxxxxxxxx
a
Suggested Warmup Time: xxx SECONDS
Module Warmup Status: xxx
b
For an example of the output data, refer to
the description on module information in
the user’s guide in the “Calibrating for
Increased Measurement Accuracy” chapter.
a. Output in hertz.
b. Can be polled in a loop to return the latest warmup time. When warmup time is finished,
the string “READY” will be returned.
N/A
<array><
L
F
>
Command Hardkey Softkey
ECALMODINF
Cal