7-4
Supplement for the Installation and Quick Start Guide (Firmware Version 7.68)
If You Encounter a Problem
If You Encounter a Problem
Power-Up Problems
Refer to Figure 2-17 to remove the fuse from the power module. You can use a continuity
light or an ohmmeter to check the fuse. An ohmmeter should read very close to zero ohms
if the fuse is good. For 115V operation, use Fuse, T 5A 125V, UL listed/CSA certified to 248
standard (part number 2110-1059). For 230V operation, use Fuse, T 4A H 250V, built to
IEC 127-2/5 standard (part number 2110-1036).