3 Dimensions and Weights
Dimensions and Weight
Site Preparation Guide 19
• Foreline pump
Standard
Oil-free (MVP-070)
Oil-free (IDP3)
21 cm (8-inch)
19 cm (7.5-inch)
18 cm (7-inch)
13 cm (5-inch)
32 cm (13-inch)
35 cm (14-incn)
31 cm (12-inch)
28 cm (11-inch)
14 cm (6-inch)
11 kg (23.1 lb)
16 kg (35.2 lb)
10 kg (21 lb)
• GC/MS operational and maintenance
access
Requires 30 cm (1 ft) to the left of the unit.
MS
7000 and 7010 Triple Quad MS
• EI Mainframe 47 cm (18.5-inch) 35 cm (14-inch) 86 cm (34-inch) 59 kg (130 lb)
• EI/CI Mainframe 47 cm (18.5-inch) 35 cm (14-inch) 86 cm (34-inch) 63.5 kg (140 lb)
• Foreline pump 28 cm (11-inch) 18 cm (7-inch) 35 cm (14-inch) 21.5 kg (47.3 lb)
• GC/MS operational and maintenance
access
Requires 30 cm (1 ft) to the left of the unit.
7200 Q-TOF MS
• Mainframe 133 cm (52.5-inch) 90 cm (34.5-inch) 100 cm (39.5-inch) 138 kg (305 lb)
• Foreline pump 28 cm (11-inch) 18 cm (7-inch) 35 cm (14-inch) 21.5 kg (47.3 lb)
7250 Q-TOF MS
• Mainframe 190 cm (74.8-inch) 90 cm (34.5-inch) 100 cm (39.5-inch) 138 kg (350 lb)
• Foreline pump DS202 28 cm (11-inch) 18 cm (7-inch) 35 cm (14-inch) 21.5 kg (47.3 lb)
• Foreline pump IDP-15 36.4 cm (14.3-inch) 33.3 cm (13.1-inch) 48.5 cm (19.1-inch) 45.5 kg (100 lb)
• GC/Q-TOF operational and maintenance
access
Requires 40 cm (16-inch) clearance on both sides of the unit.
Requires 30 cm (12-inch) clearance behind the unit.
Headspace sampler (HS)
7697A Headspace sampler
111 vial model
12 vial model
80 cm (32-inch)
61 cm (24-inch)
69 cm (27-inch)
64 cm (25-inch)
70 cm (27.5-inch)
69 cm (27-inch)
46 kg (101 lb)
38.2 kg (84 lb)
• GC with 7697A Headspace sampler Requires 69 cm (27-inch) to the right of the GC (G4557A), or 64 cm
(25-inch) to the right of the GC (G4556A).
G1888 Headspace sampler 56 cm (22-inch) 46 cm (18.1-inch) 64 cm (25-inch) 46.3 kg (102 lb)
ALS
• GC with 7693A ALS injector Requires 50 cm (19.5-inch) above the GC 3.9 kg (8.6 lb) each
• GC with 7693A ALS tray Requires 43 cm (16.8-inch) left of the GC
Requires 4.2 cm (1.7-inch) in front of GC
6.8 kg (15 lb) each
• GC with 7650A ALS injector Requires 50 cm (19.5-inch) above the GC 3.9 kg (8.6 lb) each
• GC with CTC PAL Autosampler Requires 76.6 cm (30.2-inch) above the GC and
65 to 98 cm (25.6 to 38.6-inch) to the right of the
GC, depending on configuration
Table 4 Instrument dimensions, weights, and required clearances (continued)
Product Height Width Depth Weight