4 Environmental Conditions
Heat Dissipation
Site Preparation Guide 25
Heat Dissipation
• Use Table 7 to estimate the additional heat dissipated from the equipment. Maximums
represent the heat given off when heated zones are set for maximum temperatures.
If using the optional G4522A Cooling Accessory, you will need to supply:
• A water chiller.
• Tubing and 1/8-inch Swagelok fittings to connect the chilled water and return water to the
chiller.
A container or drain to dispose of condensate from the tray.
Table 7 Heat dissipation
Instrument
8860 GC 7681 BTU/hour maximum
(8103 kJ/h)
5120 BTU/hour maximum (100 V
power option) (5402 kJ/h)
5975 Series MSD 3000 BTU/hour (3165 kJ/h)
5977 Series MSD 3000 BTU/hour (3165 kJ/h)
7010 or 7000 Triple Quad
MS
3700 BTU/hour (3904 kJ/h)
7200 or 7250 Q-TOF MS 6200 BTU/hour (6541 kJ/h)