6 GC System Power Requirements
ALS Power Requirements
38 Site Preparation Guide
ALS Power Requirements
The ALS components draw power from the GC. No other power source is required.
The G4517A controller, used with the 8860 Series GC, requires one electrical outlet with a
dedicated ground. The controller can be set for either 100-120 V or 200-240 V.
WARNING
Do not use extension cords with Agilent instruments. Extension cords normally are not rated
to carry enough power and can be a safety hazard.
The length of the power cord is 2 meters (6.6 feet).