4 Environmental Conditions
ALS Environmental Conditions
26 Site Preparation Guide
ALS Environmental Conditions
Operating the instrument within the recommended ranges optimizes instrument performance
and lifetime. The sampler system operates in the same environment as its parent GC.
See “Environmental Conditions” on page 24.
The conditions assume a non-condensing, noncorrosive atmosphere.
Table 8 Environmental conditions for operation and storage
Product
Conditions
Operating temp
range
Operating humidity range
Maximum altitude
G4513A Injector
G4514A Tray
*
G4515A Bar Code Reader*
Operation 0 to 40 °C 5–95% 4,300 m
7650 Injector Operation 0 to 40 °C 5–95% 4,300 m
G4517A Controller Operation –5 to 45 °C Maximum relative humidity of 80% for
temperatures up to 31 °C decreasing
linearly to 50% relative humidity at 40 °C
2,000 m
Storage
* The G4520A Tray with a bar code reader is available with a G4514A Tray and G4515A bar code reader.