3 Dimensions and Weights
ALS Dimensions and Weight
Site Preparation Guide 21
ALS Dimensions and Weight
Select the laboratory bench space before the system arrives. Pay special attention to the total
height requirements. Avoid bench space with overhanging shelves. See Table 5.
The instrument needs space for proper convection of heat and ventilation. Allow at least
20 cm clearance between the back of the instrument and wall to dissipate hot air.
Table 5 Required height, width, depth, and weight
Product Height (cm) Width (cm) Depth (cm) Weight (kg)
G4513A Injector 51 16.5 16.5 3.9
G4514A Tray
*
29 44 43 6.8
7650A Injector 51 22 24 4.5
Additional space requirements
GC with 7693A ALS injector 50 cm (19.5-inch) above the GC
GC with 7693A ALS tray 45 cm (17.5-inch) left of the GC
GC with 7650 ALS injector 50 cm (19.5-inch) above the GC
9 cm (3.6-inch) in front of the GC
3 cm (1.2-inch) to the left of the GC
* The G4520A Tray with a bar code reader is available with a G4514A Tray and G4515A bar code reader.

Other manuals for Agilent Technologies 8860

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8860 and is the answer not in the manual?

Agilent Technologies 8860 Specifications

General IconGeneral
BrandAgilent Technologies
Model8860
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals