6 GC System Power Requirements
Power Requirements
Site Preparation Guide 33
The GC and related equipment meet the following International Electrotechnical Commission
(IEC) classifications: Equipment Class I, Laboratory Equipment, Installation Category II, and
Pollution Degree 2.
MSD
5975 Series MSD 120
(–10% / +5%)
50/60 ± 5% 1100 (400 for foreline
pump only)
8 10 Amp
Dedicated
5975 Series MSD 220–240
(–10% / +5%)
50/60 ± 5% 1100 (400 for foreline
pump only)
8 10 Amp
Dedicated
5975 Series MSD 200
(–10% / +5%)
50/60 ± 5% 1100 (400 for foreline
pump only)
8 10 Amp
Dedicated
5977 Series MSD 120
(–10% / +5%)
50/60 ± 5% 1100 (400 for foreline
pump only)
8 10 Amp
Dedicated
5977 Series MSD 220–240
(–10% / +5%)
50/60 ± 5% 1100 (400 for foreline
pump only)
8 10 Amp
Dedicated
5977 Series MSD 200
(–10% / +5%)
50/60 ± 5% 1100 (400 for foreline
pump only)
8 10 Amp
Dedicated
MS
7010 or 7000 Triple Quad MS 120
(–10% / +5%)
50/60 ± 5% 1600 15 15 Amp
Dedicated
7010 or 7000 Triple Quad MS 220–240
(–10% / +5%)
50/60 ± 5% 1600 15 15 Amp
Dedicated
7010 or 7000 Triple Quad MS 200
(–10% / +5%)
50/60 ± 5% 1600 15 15 Amp
Dedicated
7200 or 7250 Q-TOF MS 200–240
(–10% / +5%)
50/60 ± 5% 1800 (1200 for
foreline pump)
15 15 Amp
Dedicated
HS
7697A Headspace Americas: 120
single phase
(–10% / +10%)
50/60 ± 5% 850 6.2 15 Amp
Dedicated
7697A Headspace 200/220/230/240
single/split phase
(–10% / +10%)
50/60 ± 5% 850 3.8/3.4/
3.3/3.1
10 Amp
Dedicated
All
Data system PC
(monitor, CPU, printer)
100/120
(–10% / +5%)
50/60 ± 5% 1000 10/8.3 15 Amp
Dedicated
Data system PC
(monitor, CPU, printer)
200-240
(–10% / +5%)
50/60 ± 5% 1000 4.1-5 10 Amp
Dedicated
Table 10 Power requirements (continued)
Product
Oven type
Line voltage (VAC)
Frequency
(Hz)
Maximum
continuous power
consumption (VA)
Current
rating
(amps)
Power outlet
current rating