1260 Infinity Binary LC - System User Guide 13
The Agilent 1260 Infinity Binary LC - Product Description
1
System Components
Damper and mixer can be bypassed for lowest delay volume of the binary
pump. This configuration is recommended for low flow rate applications
with steep gradients.
Figure 2 on page 13 illustrates the flow path in low delay volume mode.
For instructions on how to change between the two configurations, see the
G1312B Binary Pump User Manual.
Figure 2 The Hydraulic Path of the Binary Pump with Bypassed Damper and Mixer
NOTE
Bypassing the mixer while the damper remains in line is not a supported configuration and
may lead to undesired behavior of the binary pump.