58 1260 Infinity Binary LC - System User Guide
3 Optimization of the Agilent 1260 Infinity Binary LC
How to Achieve Higher Resolution
Data Rate
The data rate setting for the UV detector must be selected appropriately.
Selecting a data rate that is too low results in increased peak width and
lower resolution, see Figure 19 on page 58.
Figure 19 Peak width vs, UV data Rate
Table 6 on page 59 shows the dependency of peak width, resolution and
peak capacity on the data rate. For fast and ultra- fast LC, high data rates
are needed; otherwise, the separation that is achieved on the column is
destroyed in the detector.