112 Chapter 2
Instrument Functions: A - L
Display
Instrument Functions: A - L
Remote Command:
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe]?
Example: DISP:WIND:TRAC:GRAT:GRID OFF
2.5.7.2 Annotation
Turns the screen annotation on or off for all windows, however, menu key annotation will
remain on the display. The screen annotation may not be required for prints or during
remote operation.
Key Path:
Display
Factory Preset
and *RST: On
Remote Command:
:DISPlay:WINDow:ANNotation[:ALL] OFF|ON|0|1
:DISPlay:WINDow:ANNotation[:ALL]?
Example: DISP:WIND:ANN OFF
2.5.8 Display Enable (Remote Command Only)
Turns the display on/off. If enable is set to off, the display will appear to blank. This can
make the measurement run faster since the instrument does not have to update the
display after every data acquisition. There is often no need to update the display
information when using remote operation.
Sending SYSTem:PRESet or *TST commands will turn it on. *RST will not.
Key Path: None, no front-panel control.
State Saved: Not saved in instrument state.
Factory Preset: On
Factory
Default: On
Remote Command:
:DISPlay:ENABle OFF|ON|0|1
:DISPlay:ENABle?
Example: DISP:ENAB OFF

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals