Chapter 2 47
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
2.1.9 Ref Lvl Offset
Allows you to add an offset value to the displayed reference level. The reference level is the
absolute amplitude represented by the top graticule line on the display. Reference-level
offsets are entered by using the numeric keypad or programming commands. The knob and
step keys are not active.
Offsets are used when gain or loss occurs between a device under test and the analyzer
input. Thus, the signal level measured by the analyzer may be thought of as the level at
the input of an external amplitude conversion device. Entering an offset does not affect the
trace position or attenuation value, just the displayed value readouts such as reference
level and marker amplitudes.
The maximum reference level available is dependent on the reference level offset. That is,
Ref Level
Ref Level Offset must be in the range 170 to +30 dBm.
For example, the reference level value range can be initially set to values from
170 dBm
to 30 dBm with no reference level offset. If the reference level is first set to
20 dBm, then
the reference level offset can be set to values of
50 to +150 dB.
If the reference level offset is first set to
30 dB, then the reference level can be set to
values of
200 dBm to 0 dBm. In this case, the reference level is clamped at 0 dBm
because the maximum limit of +30 dBm is reached with a reference level setting of 0 dBm
with an offset of
30 dB. If instead, the reference level offset is first set to 30 dB, then the
reference level can be set to values of
140 to +60 dBm.
When a reference level offset is entered, the offset value appears on the left side of the
display under Offst (as opposed to frequency offsets which appear at the bottom of the
display.) To eliminate an offset, press
Ref Lvl Offst, 0, and dB.
Key Path:
AMPLITUDE / Y Scale
Key Notes: Only numeric entries are valid, the knob and step keys are not applicable
to this function.
Annunciation/
Annotation: The offset is displayed to left of Trace window; third from the top, just
below the scale type.
State Saved: Saved in Instrument State
Factory Preset: 0.0 dB
Range: 327.6 dB to 327.6 dB

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals