Chapter 2 49
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
2.1.11 Corrections
Accesses the Corrections menu keys that allow you to enable the corrections function and to
select which set of correction factors you wish to modify. These frequency/amplitude
corrections will be applied to the displayed data to correct for system losses/gains outside
the analyzer. Four different sets of correction data can be stored.
Key Path:
AMPLITUDE / Y Scale, More
Remote Command:
There is no equivalent remote command for this key.
2.1.11.1 Apply Corrections
Pressing
Apply Corrections (Yes) turns on the amplitude-correction factors. Corrections will
only be applied for the sets of correction factors whose correction state is set to On. To turn
a set of correction factors on, use the
Correction On Off key in the Antenna, Cable, Other, or
User menus.
Key Path:
AMPLITUDE / Y Scale, More, Corrections
Annunciation/
Annotation: When
Apply Corrections (Yes) is selected, an A will appear in the screen
annotation on the left edge of the display, whether or not a correction set
has been turned on using the
Correction (On) key in the Antenna, Cable,
Other, or User menus.
State Saved: Saved in Instrument State
Factory Preset: No
Remote Command:
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Remote Command Notes: To turn On or Off an individual correction set, use:
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]
Example: CORR:CSET:ALL ON