212 Chapter 4
Instrument Functions: P - Z
Preset
Instrument Functions: P - Z
Dependencies/
Couplings: Depends on the preset type (user, mode or factory) setting in the
System,
Power On/Preset keys.
SCPI Status Bits/
OPC Dependencies: Clears all pending OPC bits. The status byte is set to 0.
Remote Command:
:SYSTem:PRESet
Remote Command Notes: The SYSTem:PRESet command immediately presets the
instrument state to values dependent on the preset type that is currently
selected (FACTory, USER, MODE).
SYSTem:PRESet will not reset persistent functions such as IP address,
time/date display style, or auto-alignment state to their factory defaults.
Use SYSTem:PRESet:PERSistent. See Restore Sys Defaults on page 231.
SYSTem:PRESet:TYPE sets the type of preset. See Preset Type on
page 211.
Example: :SYST:PRES:TYPE MODE sets the preset mode type to mode. See Preset
Type on page 211.
:SYST:PRES presets the instrument to the currently selected preset type.
4.2.1 User Preset
This key is only available when the preset type is set to User. Press System, Power
On/Preset
, Preset Type, User.
Restores the analyzer to a user defined state that has been saved for all analyzer modes.
You defined this state when
Save User Preset was pressed. If the you have never saved a
user preset state, then the factory preset state is stored as the user preset state. If the user
preset state has been saved but the load fails for any reason, the error message: Unable to
load user state is displayed in the status line and the state is reset to whatever it was
before the
Preset key was pressed. This can sometimes happen if firmware has been
upgraded or applications have been (un)installed after the user preset state was saved.
Save User Preset can be accessed by pressing System, Power On/Preset.
Key Path:
Preset
Remote Command:
:SYSTem:PRESet See :SYSTem:PRESet on page 186.
Example: SYST:PRES:SAVE saves a user state that will be used for the preset
SYST:PRES:TYPE USER selects the user type preset
SYST:PRES immediately presets the analyzer to the user preset.

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals