Chapter 4 221
Instrument Functions: P - Z
Restart
Instrument Functions: P - Z
4.4 Restart
This function restarts a previously paused measurement at the beginning. If the
measurement is active, it will stop it as soon as possible and restart it from the
beginning.If no measurement is active and
Sweep (Single) is selected, a new sweep is
initiated.
When in
Average (On) mode, (BW/Avg, Average) the averaging function is restarted (the
trace is reset and the average number is reset to zero).
Key Path: Front-panel key. It can also be found under
Meas Control.
Remote Command:
:INITiate:RESTart
Remote Command Notes: This command is equivalent to sending an :ABORt command
followed by an :INITiate[:IMMediate] command. See “Abort the Sweep
or Measurement (Remote Command Only)” on page 194. for more
information.
Example: INIT:REST