EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E4440A

Agilent Technologies E4440A User Manual

Agilent Technologies E4440A
434 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #221 background imageLoading...
Page #221 background image
Chapter 4 221
Instrument Functions: P - Z
Restart
Instrument Functions: P - Z
4.4 Restart
This function restarts a previously paused measurement at the beginning. If the
measurement is active, it will stop it as soon as possible and restart it from the
beginning.If no measurement is active and
Sweep (Single) is selected, a new sweep is
initiated.
When in
Average (On) mode, (BW/Avg, Average) the averaging function is restarted (the
trace is reset and the average number is reset to zero).
Key Path: Front-panel key. It can also be found under
Meas Control.
Remote Command:
:INITiate:RESTart
Remote Command Notes: This command is equivalent to sending an :ABORt command
followed by an :INITiate[:IMMediate] command. See Abort the Sweep
or Measurement (Remote Command Only) on page 194. for more
information.
Example: INIT:REST

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals