Chapter 2 35
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
Remote Command:
[:SENSe]:POWer[:RF]:ATTenuation <rel_power>
[:SENSe]:POWer[:RF]:ATTenuation?
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Remote Command Notes: . The Reference Level setting may be affected when the
Attenuation is changed. See
Ref Level.
Example:. POW:ATT 30
POW:ATT?
POW:ATT:AUTO ON
POW:ATT:AUTO?
2.1.3 Scale/Div
Sets the units per vertical graticule division on the display. This function is only available
when
Scale Type (Log) is selected and the vertical scale is power, or Demod View is on and
the vertical scale is hertz. When
Scale Type (Lin) is selected, Scale/Div is grayed out.
Key Path:
AMPLITUDE / Y Scale
State Saved: Saved in Instrument State
Factory Preset: 10 dB, in logarithmic scale type
Fundamental
Units: dB; kHz
Terminators: dB; Hz, kHz, MHz, GHz
Default
Terminator: dB; kHz
Range: 0.1 dB to 20 dB
1 kHz to 240 kHz (When
Demod View (On) is selected.)
Remote Command:
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:PDIVision <power>
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:PDIVision?
Example: DISP:WIND:TRAC:Y:PDIV 5 DB

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals