Chapter 4 263
Instrument Functions: P - Z
System
Instrument Functions: P - Z
Key Path: System, Reference
State Saved: Not Saved in Instrument State. Neither the external reference frequency
nor the state of this function (Int or Ext) are affected by factory preset or
power cycle. Reset to the factory default (Int, 10 MHz) by pressing
System,
Restore Sys Defaults.
Default: Internal, 10 MHz
Range: 1 MHz to 30 MHz
Remote Command:
[:SENSe]:ROSCillator:SOURce INTernal|EXTernal
[:SENSe]:ROSCillator:SOURce?
[:SENSe]:ROSCillator:EXTernal:FREQuency <value>
[:SENSe]:ROSCillator:EXTernal:FREQuency?
Example: Before switching to the external reference source, specify the frequency of
the external reference that you plan to use.
ROSC:EXT:FREQ 20 MHz sets the external reference frequency to 20 MHz,
but does not select the external reference.
ROSC:SOUR EXT selects the external reference.
4.9.6.2 10 MHz Out
Switches the 10 MHz out signal on the rear panel of the analyzer on or off.
Key Path: System, Reference
State Saved: Not Saved in Instrument State. Not affected by factory preset or power
cycle. Reset to the factory default (Off, 10 MHz) by pressing
System, Restore
Sys Defaults
.
Remote Command:
[:SENSe]:ROSCillator:OUTPut[:STATe] OFF|ON|0|1
[:SENSe]:ROSCillator:OUTPut[:STATe]?
Example: ROSC:OUTP ON