Chapter 6 357
Using the STATus System
Using the Status Registers
Using the STATus System
To query the status byte register, send the command *STB? The
response will be the decimal sum of the bits which are set to 1. For
example, if bit number 7 and bit number 3 are set to 1, the decimal sum
of the 2 bits is 128 plus 8. So the decimal value 136 is returned. The
*STB command does not clear the status register
In addition to the status byte register, the status byte group also
contains the service request enable register. This register lets you
choose which bits in the status byte register will trigger a service
request.
Send the *SRE <number> command where <number> is the sum of the
decimal values of the bits you want to enable plus the decimal value of
bit 6. For example, assume that you want to enable bit 7 so that
whenever the standard operation status register summary bit is set to 1
it will trigger a service request. Send the command *SRE 192 (because
128 + 64). You must always add 64 (the numeric value of RQS bit 6) to
your numeric sum when you enable any bits for a service request. The
command *SRE? returns the decimal value of the sum of the bits
previously enabled with the *SRE <number> command.
The service request enable register presets to zeros (0).

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals