362 Chapter 6
Using the STATus System
Using the Status Registers
Using the STATus System
Questionable Status Register
The questionable status register monitors the instruments condition to
see if anything questionable has happened to it. It is looking for
anything that might cause an error or a bad measurement like a
hardware problem, an out of calibration situation, or a unusual signal.
All the bits are summary bits from lower-level event registers.
Bit Condition Operation
3 Power summary The instrument hardware has detected a
power unleveled condition.
4Temperature
summary
The instrument is still warming up.
5 Frequency
summary
The instrument hardware has detected an
unlocked condition or a problem with the
external frequency reference.
8 Calibration
summary
The instrument has detected a hardware
problem while doing the automatic internal
alignment process.
9Integrity
summary
The instrument has detected a questionable
measurement condition such as: bad timing,
bad signal/data, timeout problem, signal
overload, or meas uncal.

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals