Chapter 2 77
Instrument Functions: A - L
BW/Avg
Instrument Functions: A - L
Key Path: BW/Avg, Avg/VBW Type
or Auto Couple, Avg/VBW Type
Remote Command:
See Avg/VBW Type on page 74.
Example: AVER:TYPE RMS
2.3.5.4 Voltage Avg
In this Average type, all filtering and averaging processes work on the voltage of the
envelope of the signal. This scale is good for observing rise and fall behavior of AM or
pulse-modulated signals such as radar and TDMA transmitters, but its response to
noise-like signals is 1.049 dB lower than the average power of those noise signals. This is
compensated for in the Marker Noise function. When this type of averaging is selected,
VAvg appears on the left side of the display.
In the equation for averaging on this scale (below), N is the number of averages
accumulated. (In continuous sweep mode, once N has reached the Average Number, N
stays at that value.)
Assumes all values in dB.
Key Path:
BW/Avg, Avg/VBW Type
or Auto Couple, Avg/VBW Type
Remote Command:
See Avg/VBW Type on page 74.
Example: AVER:TYPE SCAL
New Avg 20
N1()10×
Old Avg
20
--------------------
10
New data
20
----------------------
+
N
------------------------------------------------------------------------------




log×=

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals