Chapter 2 81
Instrument Functions: A - L
Det/Demod
Instrument Functions: A - L
Remote Command:
[:SENSe]:DETector[:FUNCtion] NORMal|AVERage|POSitive|SAMPle|NEGative|QPEak
|EAVerage|EPOSitive|MPOSitiv|RMS
[:SENSe]:DETector[:FUNCtion]?
The query returns a name that corresponds to the detector mode as shown by the following
terms:
Remote Command Notes:
The query returns a name that corresponds to the detector mode. The RMS
selection is an alias which selects the Average detector and Power
Averaging. Therefore, if RMS has been selected, the query will return the
AVER string.
Example: DET POS
2.4.1.1 Auto
The system selects normal detection as the default, but if a condition arises where a
different type of detection scheme would be better utilized, the system uses the alternate
scheme. For example, when in Auto mode, the Marker Noise function uses Average
detection because the system determines that the average detector has better
speed/variance tradeoffs for noise-type signals.
Refer to Figure 2-1, which shows a decision tree of how detection type is determined.
NORMal Normal
AVERage Average
POSitive Peak
SAMPle Sample
NEGative Negative peak
QPEak Quasi Peak
EAVerage EMI Average
EPOSitive EMI Peak
MPOSitive MIL Peak
RMS RMS (alias)

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals