Agilent 1260 Infinity II Preparative Autosampler User Manual 115
Troubleshooting and Diagnostics
7
Maintenance Functions
To prepare autosampler for transportation
The park arm function moves the gripper and transport slider to the home
position. Reinstall the transport foam to secure the instrument against a
mechanical stop. The autosampler can be switched OFF after the preparation.
1 Select Park Arm.
2 When the arm is in the park position, install Transport-Protection
(G7129-40050) . Then the autosampler is ready for shipment, and can be
switched OFF.
When Before transporting or shipping the module.
CAUTION
Unsecured Transportation of the module
Unsecured transportation of the module may result in mechanical damage to the
gripper and transport slider.
Always secure the arm in the park position plus transport foam.
NOTE
Before parking the gripper arm, ensure there is no vial in the gripper. Use the Release
Gripper function to remove the vial.

Table of Contents

Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals