132 Agilent 1260 Infinity II Preparative Autosampler User Manual
8
Error Information
General Error Messages
Leak Sensor
Error ID: 0064
Leak
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the main board.
Probable cause Suggested actions
1
Loose fittings.
Ensure all fittings are tight.
2
Broken capillary.
Exchange defective capillaries.
3
Leaking rotor seal or needle seat.
Exchange the rotor seal or seat capillary.
4
Defective metering seal.
• Exchange the metering seal.
• Make sure the leak sensor is thoroughly dry
before restarting the autosampler.
5
Leaking peristaltic pump
Exchange the peristaltic pump.
The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf
respectively.