132 Agilent 1260 Infinity II Preparative Autosampler User Manual
8
Error Information
General Error Messages
Leak Sensor
Error ID: 0064
Leak
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the main board.
Probable cause Suggested actions
1
Loose fittings.
Ensure all fittings are tight.
2
Broken capillary.
Exchange defective capillaries.
3
Leaking rotor seal or needle seat.
Exchange the rotor seal or seat capillary.
4
Defective metering seal.
Exchange the metering seal.
Make sure the leak sensor is thoroughly dry
before restarting the autosampler.
5
Leaking peristaltic pump
Exchange the peristaltic pump.
NOTE
The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf
respectively.

Table of Contents

Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals