22 Agilent 1260 Infinity II Preparative Autosampler User Manual
1
Introduction
Leak and Waste Handling
Leak and Waste Handling
Leak Sensor
Leak Sensor
Leak Sensor
1
NOTE
The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf
respectively.
CAUTION
Solvent incompatibility
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor. The
material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
Do not use DMF.
NOTE
The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf
respectively.